Fakeeh Logo

Fakeeh Hospital Makkah

Medical Equipment Technical Requirements &
Utility Consumption Report

Project: CSSD Renovation

Date: November 2025

Fakeeh Hospital Makkah Building
Logo

Utility Requirements & Consumption Summary

Parameter Sterilizer 1 (Small) Sterilizer 2 (Large) Washer Disinfector Strict Requirement (Line Sizing)
Cold / Feed Water Supply
Pressure 3 - 5 bar 3 - 5 bar 2 - 8 bar 3 - 5 bar (Regulated)
Flow Rate (Capacity) 0.9 m³/h 0.9 m³/h ≥ 15 L/min (0.9 m³/h) ≥ 0.9 m³/h per machine
Consumption (Avg) ~0.09 m³/h ~0.13 m³/h Varies by cycle Max 0.13 m³/h load
Hardness / Temp 1-13 °d / <15°C 1-13 °d / <15°C 0-30 °dH / 5-30°C 1-13 °dH @ <15°C
Demineralized / RO Water
Flow Rate (Capacity) 0.05 m³/h 0.25 m³/h (Implied by pressure) ≥ 0.25 m³/h per machine
Conductivity < 15 µS/cm < 15 µS/cm 3 - 200 µS/cm < 15 µS/cm (Target <10)
Electrical & Environmental
Power Supply 400V 3NPE, 50Hz 400V 3NPE, 50Hz 400V 3NPE 400V 3NPE ±10%
Connected Load 17 kW (25A Fuse) 24.5 kW (36A Fuse) (Refer to Manual) Min 36A / 25kW Supply
Heat Emission 2.83 kW 3.63 kW ~1.5 kW HVAC Load ~3.63 kW
Cooling Water (Opt.) 0.8 m³/h Flow 1.2 m³/h Flow N/A Line: 1.2 m³/h (if used)
```