Regular Maintenance & Monitoring Responsibility Schedule

Reverse Osmosis Dialysis Water Treatment Unit — HP5500D 1000 P · 60 Hz · Hercotherm G 18 kW · DIATRON 5500‑II
ManufacturerHerco Wassertechnik GmbH
Project / CustomerEzdihar · United Medical Supplies Ltd.
Order / Doc No.AF20‑008099
Item No.465 916
SourceHP5500D 1000 P 60 Hz Manual (BA_KSA)
PreparedOcean Gates Trading Est.
Source note. This schedule is compiled from the operating manual chapters that were machine‑readable (A – General & Safety, C – Technical Data / Product Description, E – Commissioning / Decommissioning, F – Operation / Monitoring, G – Troubleshooting) and the manual's Wear & Spare Parts list. The manual's dedicated Chapter H “Maintenance / Disinfection” periodic table could not be auto‑extracted from the 21 MB combined PDF (the reader stopped just before it). The rows below are all sourced & chapter‑referenced; the exact Chapter H interval table can be transcribed in verbatim once those pages are supplied.
Responsible party (per manual roles / password levels, Ch. A §2.5–2.6 & Ch. F):
Operator / User Daily operation, monitoring, alarm acknowledgement, hot‑sanitisation start, week‑timer (User pw 1234)
Technician Logging, diagnosis, calibration, communication, HS sequences (Technician pw)
Service / Manufacturer Chemical disinfection, default settings, part replacement, repairs (Service pw / Herco)
Plant Operator (org.) Ensures qualified staff, training, records, safe isolation
Alarm category (Ch. G):
Cat I Warning — yellow, unit keeps running
Cat II Fault — red, unit keeps running
Cat III Fault — red, emergency shutdown

1 · Continuous Monitoring — each measured valueCh. F §4–6 · Ch. G §1.4

The DIATRON 5500‑II monitors every value continuously and logs it (SD card / USB). The Operator watches the display and acknowledges alarms; Cat II/III faults or anything needing a repair are escalated to qualified technical / service staff.

Monitored valueTag Alarm / trigger & category ActionResponsible
Feed‑water conductivityQ01 FC010 sensor out of range Acknowledge; check probeOperator
Permeate conductivity RO pass 1Q02 FC012 > P063 warning · FC013 > P064 fault · FC011 sensor fail Acknowledge; persistent → check membranes / calibrate OperatorTechnician
Permeate conductivity loop supplyQ03 FC015 warning · FC016 > P084 fault · FC014 < P082 Cat III → emergency shutdown; manual reset after cause removed OperatorService
Feed‑water temperatureT01 FC018 > P033 warning · FC017 sensor fail Acknowledge; check feed / heatingOperator
Permeate temp RO pass 1T02 FC020 warning · FC021 fault · FC019 sensor fail Acknowledge; investigate over‑temperatureOperator
Permeate temp loop supplyT03 FC023 warning · FC024 fault (→ recirc. until OK) · FC022 sensor fail Acknowledge; unit recirculates until temp OKOperator
Loop & hot‑san. tempsT04–T09, T11, T12 FC025–FC033 sensor / over‑temp → hot sanitisation aborted, cool‑down Check heaters / sensors; re‑run HS Technician
Fine‑filter differentialPr01–Pr02 FC044 “Check fine filter F1” > P183 Replace fine‑filter cartridge F1 Operator / Technician
Operating pressure pass 1Pr05 / Pr06 FC046 / FC047 too high Manual reset after cause removed; check valves / membranes Technician
Operating pressure pass 2Pr08 / Pr09 FC048 / FC049 too high · FC043 low pressure pass 2 Manual reset; check pump P2 / valves Technician
Permeate loop pressurePr04 / Pr10 FC045 low pressure loop · FC042 permeate over‑pressure Check loop / overflow valve DV02 setting Technician
Flow sensors (all)Fl01–Fl05 FC004–FC008 no flow pulses (300 s) Acknowledge; check paddle‑wheel sensorOperator
Loop leakage detectionFl01–Fl02 Δ FC009 difference > P015 → shutdown Locate & stop leak; manual reset OperatorService
Holding‑tank levelB1 FC040 tank empty (shutdown) · FC041 tank full Check feed supply / level switch Operator
Valve position transmittersG02 (V10), G03 (V11) FC002 / FC003 position ≠ operating mode Correct valve position / sensor gap Technician

2 · Routine Operating ChecksCh. F §2 · G §1.1

TaskIntervalResponsibleRef.
After dialysis, switch the unit to STANDBY (not OFF) to prevent microbiological contamination during idle time Every session / dailyOperatorCh. F §2.3
Keep holding‑tank B1 cover closed; never drink / use permeate for drinks ContinuousOperatorCh. F §2
Review current operating values (temperature, conductivity, feed/return flow, consumption) DailyOperatorCh. F §5
Ensure SD card is inserted; back up log data to USB PeriodicOperator / TechnicianCh. F §4
Keep operation & service records available (needed by manufacturer for support) Rolling — last 4 monthsPlant OperatorCh. G §1.1

3 · Disinfection, Hot Sanitisation & PreservationCh. C §4.2 · E · F

TaskInterval / triggerResponsibleRef.
Chemical disinfection — ~0.2% peracetic acid; ~30 min recirculation + ~2 h hold + flush; confirm disinfectant‑free before return to service Regularly, per site hygiene plan (idle‑time only). Prevents deposits & microbial growth Service (service pw) + Operator (PPE / chemical handling)Ch. C §4.2 · F pw‑levels
PPE when handling peracetic acid / preservative — eye protection, rubber gloves, apron; remove spills immediately Whenever chemicals handledOperatorCh. A §2.7 · E §2.2
Hot sanitisation up to 90–95 °C (Hercotherm G 18 kW) via week timer — programmes HS1–HS4 Scheduled (auto week timer), per hygiene planOperator starts (User pw)Ch. F §3 · C
Preserve the unit (sodium bisulphite + glycerine) when idle If idle > 7 daysService / TechnicianCh. E §2.1
Replace preservation agent; flush & disinfect before re‑commissioning After 12 months at the latestService / TechnicianCh. E §2.1

4 · Pre‑treatment & Feed‑water CareCh. A §2.11 · C · D

TaskIntervalResponsibleRef.
Check activated‑carbon filter; replace when exhausted (dispose per local rules) At regular intervalsOperator / TechnicianCh. A §2.11
Verify feed water stays within spec: softened, particle‑free, free chlorine not detectable, max TDS 1000 mg/l, SDI₁₅ ≤ 3, 5–25 °C, 1.5–6 bar OngoingOperatorCh. C §2
Replace fine‑filter cartridge F1 on differential‑pressure alarm (FC044) On alarm / condition‑basedOperator / TechnicianCh. G §1.4

5 · Wear & Spare‑Part Replacement IntervalsManual: Wear & Spare Parts list (AF20‑008099)

Replacement / service basis stated in the manual: 5,000 operating h/yr (RO), 150 hot disinfections/yr (RO), 50 hot disinfections/yr (loop). Items marked S are carried on the service vehicle. Replacement is a Service / Manufacturer task.

ComponentTag(s)Replacement intervalResponsible
Membrane modules (LOW4 / ESPA4)X1.1‑1.3 / X2.1‑2.34–8 yearsService
Sets of sealingsV14, V03, V05…30,000 h or 10 years / as neededService
Sealing set (hot‑dis. duty)1,000 hot dis. or 10 years / as neededService
Pump shaft seals CRN 3‑25 / CRN 3‑19P1 / P26 yr / 2–5 yrService
Pump seal set CM‑G 3‑3P36 yearsService
Inline flow sensor 3SPL01 / L02 / L0315,000 h or 3 yearsService
Paddle‑wheel flow fittingsFl01–Fl0530,000 h / as neededService
Control‑cabinet ventilation filter mats1M21 yearTechnician
Conductivity measuring cells (ZK1.0 / ZK0.1)Q01 / Q02 / Q03As neededService
Motor contactors 5.5 kW / 7.5 kW1K1,1K2 / 2K2A,2K2B200 hot dis. or 3 yearsService
DIATRON 5500‑II CPU/Display, basic & extension modulesA0 / A1 / A26 yearsService
Solenoid / ball / needle / sampling valves, overflow & check valvesV01, V09, DV01–DV04…As neededService
Motor‑protection switches, circuit breakers, fuses, relays1Q1‑1Q3, 2F2…As neededService
Hercotherm heater / transmitter TMT127 / bimetal / PT100W2A/W2B, T04/T057‑day lead / as neededService

6 · Responsibility & Safety Framework for MaintenanceCh. A §2.5, 2.6, 2.10, 2.12

DutyResponsibleRef.
Define scope, competence & supervision of personnel; train staff who lack the required knowledgePlant OperatorCh. A §2.6
Ensure operation, maintenance, inspection & assembly are done only by suitably qualified staff (medical tasks: doctor / dialysis nurse)Plant OperatorCh. A §2.5
Before repair/maintenance: take unit out of service, secure against restart, de‑energise & depressurise, wear PPE, reinstate all guards afterwardsService / TechnicianCh. A §2.10
Use only original / manufacturer‑approved spare parts; no unauthorised modification or controller patch (else warranty & conformity void)Service / ManufacturerCh. A §2.12
Emergency control mode (unmonitored conductivity) only after consulting the doctor in chargeOperator + doctorCh. F §2