According to EN 60204-1, EN 62353
| Check | Status | Remark |
|---|---|---|
| 1.1 Blind plugs + cabinet key present | β OK | |
| 1.2 Covers for main terminals present | β OK | |
| 1.3 Cabinet labelled with lightning arrow | β OK | |
| 1.4 Type plate present | β OK | |
| 1.5 Wire colours according to wiring diagram | β OK | |
| 1.6 IP rating maintained | β OK | |
| 1.7 Cover / insulation of live parts | β OK | |
| 1.8 Cover & marking of mains terminals | β OK | |
| 1.9 Separation of N and PE terminals | β OK | |
| 1.10 Cable & fuse sizing per wiring diagram | β OK | |
| 1.11 PE sticker at protective conductor terminals | β OK | |
| 1.12 Earthing of control circuits | β OK | |
| 1.13 Earthing of cabinet, door, frame | β OK | |
| 1.14 Sticker "Screw contacts tightened" present | β OK | |
| 1.15 Screw contacts tightened | β OK | |
| 1.16 Functional test of fuses / motor protectors | β OK | Motor protector: 1gch SLP |
| Measurement | Target | Tolerance | Actual |
|---|---|---|---|
| 1.20 Protective conductor continuity | < 0.3 Ξ© | β | 0.060 Ξ© |
| 1.21 Insulation resistance | > 2 MΞ© | β | > 3.10 MΞ© |
| 1.22 Equipment leakage current | < 5 mA | β | 0.028 mA |
| 1.24 Residual voltage protection | 1s < 60V | β | 9 V |
According to EN 60204-1, EN 62353 β Additional checks for hot sanitization
| Check | Status |
|---|---|
| 1.1 Blind plugs + cabinet key | β OK |
| 1.2 Covers for main terminals | β OK |
| 1.3 Lightning arrow marking | β OK |
| 1.4 Type plate | β OK |
| 1.5 Wire colours per diagram | β OK |
| 1.6 IP rating | β OK |
| 1.7 Cover/insulation of live parts | β OK |
| 1.8 Cover & marking mains terminals | β OK |
| 1.9 Separation N / PE | β OK |
| 1.10 Cable/fuse sizing per diagram | β OK |
| 1.11 PE stickers | β OK |
| 1.12 Earthing control circuits | β OK |
| 1.13 Earthing cabinet/door/frame | β OK |
| 1.24 Functional test of fuses | β OK |
| Measurement | Target | Tolerance | Actual |
|---|---|---|---|
| 1.20 Protective conductor continuity | < 0.3 Ξ© | β | 0.085 Ξ© |
| 1.21 Insulation resistance | > 2 MΞ© | β | 134 MΞ© |
| 1.22 Equipment leakage current | < 5 mA | β | 0.0002 mA |
| 1.24 Residual voltage | 1s < 60V | β | 9 V |
| Check | Status |
|---|---|
| 2.1 Motor protector setting per wiring diagram | β OK |
| 2.2 Jumper settings Diatron 5500 | β OK |
| 2.4 Parameter settings Diatron 5500 per parameter list | β OK |
| 2.5 SD card inserted and formatted | β OK |
| 2.6 Remote monitoring PC parameterization per checklist | β OK |
| 2.7 Visualization parameterization per checklist | β OK |
| 2.8 Teamviewer ID entered in circuit diagram | β OK |
| Check | Status |
|---|---|
| 3.1 LF and temperature measurements calibrated/checked | β OK |
| 3.3 Sensor fault simulation Q02 / T02 | β OK |
| 3.4 Sensor fault simulation Q03 / T03 | β OK |
| 3.5 Watchdog simulation | β OK |
| 3.8 W2 shutdown trigger | β OK |
| 3.15 Empty alarm simulation (transfer tank) | β OK |
| 3.16 Hard water alarm simulation | β OK |
| 3.17 Position error simulation G02, G03 (normal mode) | β OK |
| 3.18 Moisture detector simulation | β OK |
| 3.19 Ring line leakage | β OK |
| 3.20 Heater β temperature sensor assignment | β OK |
| 3.21 Subloop check (if present) | β OK |
| 3.22 Universal digital I/O check | β OK |
| 3.23 Measure all heater currents during hot sanitization | β OK |
| 3.24 Check 230VAC at all power terminals of additional/intermediate heaters during hot sanitization | β OK |
| Check | Target | Tolerance | Actual |
|---|---|---|---|
| 4.1 Setting DV02 | 6.5 bar | Β±0.2 | 6.6 |
| 4.2 Setting DV01 (permeate basic setting without reduction) | 1.5 bar | +1.0 | 2.4 |
| 4.4 Feed water quantity transfer tank (orifice at feed inlet after V01) | 1600 l/h @ 2 bar | +200 / +0.5 | 1600 l/h / 2.5 l/s |
| 4.5 Concentrate control valve stage 1 (V17) normal mode | 50% of 6.7 | Β±10% | 200 l/h |
| 4.6 Concentrate control valve stage 2 (V15) normal mode | 50% of 6.15 | Β±10% | -120 l/h |
| 4.7 Pressure maintenance in ring line (Pr04) at system stop | 0.2 bar | +1.0 | 0.64 |
| 4.8 Setting DV03 (concentrate amount constant regardless of V09 state) | per 6.7 | Β±10% | 380 / 400 |
| Check | Status |
|---|---|
| 5.1 Normal mode manual/automatic | β OK |
| 5.2 Temperature rejection V09 | β OK |
| 5.3 LF-dependent concentrate rejection via V09 | β OK |
| 5.4 Temperature rejection V03 | β OK |
| 5.5 Winter mode | β |
| 5.6 Emergency mode level 1 | β OK |
| 5.7 Emergency mode level 2 | β OK |
| 5.8 Emergency control mode | β OK |
| 5.10 Hot sanitization: ring + devices | β OK |
| 5.11 USB interface function: protocol saved to USB stick | β OK |
| Measurement | Target | Tolerance | Actual |
|---|---|---|---|
| 6.1 Inlet pressure | > 2.0 bar | β | 3.5 |
| 6.2 Raw water temperature | 10β35 Β°C | β | 11.3 |
| 6.3 Raw water conductivity | 100β1600 Β΅S/cm | β | 36.1 |
| 6.4 Conductivity transfer tank (concentrated) Q01 | 1600β2000 Β΅S/cm | per test | 1500 |
| 6.5.1 Operating pressure (1PR05) | max 22 bar | β | 11.8 |
| 6.6 Concentrate pressure (PR06) | max 22 bar | β | 10.2 |
| 6.7 Concentrate flow (FL03) | 390 l/h | Β±10% | 350 |
| 6.8 Permeate flow (FL04) | 1180 l/h | Β±10% | 1180 |
| 6.9 Permeate conductivity (Q02) | 3% of 6.4 | β | 13.5 Β΅S/cm |
| 6.10 Permeate temperature (T02) | 10β35 Β°C | β | 24.5 |
| 6.15 Return pressure ring line (1Pr04) | 1.5β2 bar | +1.0 bar | 32.4 |
| Measurement | Target | Tolerance | Actual |
|---|---|---|---|
| 6.11 Supply pressure (PR07) | > 2.0 bar | β | 3.2 |
| 6.12.1 Operating pressure (PR08) | max 16 bar | β | 10.4 |
| 6.12.2 Operating pressure (PR08) at HR | < 3 bar | β | β |
| 6.13 Concentrate pressure (PR09) | max 16 bar | β | 9.0 |
| 6.14 Permeate back pressure (PR10) | 4 bar | Β±10% | 4.1 |
| 6.15 Concentrate flow (FL05) normal/emergency | 180 / 330 | Β±10% | 190 |
| 6.16 Permeate flow (FL01) normal/emergency | 1000 l/h | Β±10% | 995 |
| 6.17 Permeate conductivity (Q03) | 1.5% of 6.4 | β | 11.5 Β΅S/cm |
| 6.18 Permeate temperature (T03) | 10β35 Β°C | β | 25.2 |
| Check | Status |
|---|---|
| 7.1 Process engineering check per R+1 scheme | β OK |
| 7.2 Flow sensors checked for correct nominal size | β OK |
| 7.3 Labeling per P&ID + plate replacement + positions | β OK |
| 7.4 Correctness of labeling language | β OK |
| 7.5 Copy of test protocol, parameter list, wiring diagram in cabinet | β OK |
| 7.6 Cable labeling present | β OK |
| 7.7 Parameter backup and test protocol saved | β OK |
| 7.8 Distance between inductive sensor G02/G03 and ball valve < 1mm | β OK |
| 7.9 Analog output check per wiring diagram | β |
| 7.7 (dup) Pump rotation direction | β OK |
| 7.8 (dup) Component tightness | β OK |
| 7.11 "Caution Hot" sticker on ring inlet, after cover installation | β OK |
| 7.12 Extra undrilled blank loose on inlet with plastic bag | β OK |
| 7.13 Thermal image of heater contactors taken (during HR) | β OK |
| 7.14 Ferrites installed on controller | β OK |
| 7.15 Measuring cell Q01 serial number | 02822 89 70102 03 800 49 |
| Parameter | Description | Setpoint |
|---|---|---|
| P055 | T02 fault Β°C | 35 |
| P062 | Q02 < range Β΅S/cm | 1 |
| P063 | Q02 warning Β΅S/cm | 35 |
| P064 | Q02 fault Β΅S/cm | 40 |
| P072 | T03 < range Β°C | 2 |
| P073 | T03 V09 to drain Β°C | 25 |
| P074 | T03 warning Β°C | 30 |
| P075 | T03 fault Β°C | 35 |
| P082 | Q03 < range Β΅S/cm | 1 |
| P083 | Q03 warning Β΅S/cm | 25 |
| P084 | Q03 fault Β΅S/cm | 30 |
| P092 | T04 < range Β°C | 2 |
| P093 | T04 end heat up Β°C | 85 (80) |
| P094 | T04 limit max Β°C | 95 |
| P095 | T04 > range Β°C | 99.9 |
| P100βP104 | W2/T05 limits | range 0β100, off 90, hyst. 2 |
| P110βP114 | W3/T06 limits | range 0β100, off 90, hyst. 2 |
| P120βP124 | W4/T07 limits | range 0β100, off 90, hyst. 2 |
| P130βP134 | W5/T08 limits | range 0β100, off 90, hyst. 2 |
| P140βP144 | W6/T09 limits | range 0β100, off 90, hyst. 2 |
| P150βP154 | W1/T11 limits | range 0β100, off 90, hyst. 2 |
| P160βP163 | W12/T12 limits | range 0β100, off 90 |
| Parameter | Description | Setpoint |
|---|---|---|
| HS116 | Standby after HS | 30 min |
| D010 | t-recirculation | 30 min |
| D011 | t-residence | 2 h |
| D012 | t-rinse | 2 h |
| D013 | t-impulse RO valves | 20 s |
| D014 | t-interval RO valves | 3 min |
| D015 | t-impulse subloop valves | 60 s |
| Terminal | Signal | Type | Function |
|---|---|---|---|
| X3108-110 | PR04 | AI11 | Pressure return ring line |
| X3111-113 | PR05 | AI12 | Operating pressure |
| X3114-116 | PR06 | AI13 | Concentrate pressure |
| X3117-119 | PR10 | AI14 | Permeate back pressure |
| X3120-122 | PR07 | AI15 | Supply pressure |
| X3123-125 | PR08 | AI16 | Operating pressure 2 |
| X3126-128 | PR09 | AI17 | Concentrate pressure 2 |
| X3129-131 | T05 | AI18 | Temperature 5 |
| X3132-134 | AE | AI19 | Analog input (reserved) |
| X3135-137 | AE | AI20 | Analog input (reserved) |
| X251-53 | Uni-AAA001 | DI | Universal digital input |
| X386-88 | A002 | DI | Digital input |
| X389-91 | P2 | AI | Analog input |
| X392-94 | AO | AO | Analog output |
| X51-53 | DI30βDI31 | DI | Digital inputs (disabled) |
π Note: X2 = Mainboard, X3 = Expansion board, 0X.. = VBS (Valve Bus System). Some inputs marked "disabled" are not used.
Created by: Lei RO
Date: 01.03.21
Checked / Approved: (signature)
Date: 18.03.21
Temp: 21.8 Β°C
Humidity: 8:0.95
This protocol is a translation of the original German document (AF20-008099_0A465916).
All measurements and checkmarks correspond to the original test record.